9. SALINITY

9.1. PREPARATION

Pour small quantities of MA9066 salinity calibration solution
into clean beakers. To minimize cross-contamination, use two
beakers: one for rinsing the probe and the other for calibration.
Note: When the meter is switched on, it starts measuring with the
previously selected range (conductivity, TDS or salinity).
Note: A new EC calibration automatically clears the %NaCl
calibration . "NO CAL" message is displayed.

9.2. CALIBRATION

Press RANGE/ to select Salinity mode. %NaCl tag is displayed.
%NaCl calibration is a one-point calibration at 100.0% NaCl.
Place the probe in the calibration solution making sure that the
sleeve holes are completely submersed. Center the probe away
from the bottom or beaker walls.
Lift and lower the probe to refill the center cavity and tap the
probe repeatedly to remove any air bubbles that may have been
trapped inside the sleeve.
Press CAL/EDIT to enter to calibration mode. The CAL tag is
displayed at the top of the screen. The first LCD line displays the
NaCl reading and the third LCD line the closest calibration point.
The hourglass symbol and "WAIT" message (blinking) are
displayed until the reading is stable.
When the reading is stable and close to the selected calibration
standard, "SOL STD" message and ACCEPT tag (blinking) are
displayed. Press GLP/ACCEPT key to confirm calibration.
The instrument displays "SAVING", stores the calibration values
and returns to Measurement mode.