EC Range
Options: AUTO (default), 29.99 S/cm, 299.9 S/cm, 2999 S/cm,
29.99 mS/cm, 200.0 mS/cm
Note: Absolute conductivity — up to 500.0 mS/cm — is the
conductivity value without temperature compensation.
With the probe connected, use keys to change the value.
When autoranging, the meter automatically chooses the optimum
conductivity range to maintain the highest possible accuracy.
Note: Selected EC range is active during measurements only.
If exceeded, the full-scale value is displayed blinking. Logged
data is displayed in S/cm in the CSV files.
TDS Range
Options: AUTO (default), 14.99 mg/L, 149.9 mg/L, 1499 mg/L,
14.99 g/L, 100.0 g/L
Note: Absolute TDS — up to 400.0 g/L (with 0.8 factor) — is the
TDS value without temperature compensation.
With the probe connected, use keys to change the value.
When autoranging, the meter automatically chooses the optimum
TDS range to maintain the highest possible accuracy.
SETUP 17