Milwaukee MW306 MAX Instruction Manual page 27

Ec / tds / nacl / temperature portable meter
Table of Contents

Advertisement

Use the  keys to view the current temperature coe icient.
The value is displayed along with the Cell Factor (C.F.) on the
third LCD line.
To change the temperature coe icient, see SETUP section
for details.
A temperature coe icient must also be set for the sample.
Note: If the reading outside of range when the range is set to
automatic, the full-scale value (200.0 mS/cm for MTC/ATC or
500.0 mS/cm for No TC) is displayed blinking.
y Manual (MTC): The temperature value, shown on the second
LCD line, can be manually set using the  keys. When in
MTC mode, the °C tag is displayed blinking.
y No Temperature Compensation (NO TC): The temperature
value is displayed, but not taken into account. When this option
is selected, the NO TC tag is displayed. The reading displayed
on the first LCD line is the uncompensated EC or TDS value.
Note: Temperature-compensation and absolute conductivity
(NO TC) are configured in Setup.
TDS Measurement
Press RANGE/ to select TDS range.
y The TDS reading is displayed on the first LCD line and the
temperature reading on the second LCD line.
y Measured value is displayed in the set parameter unit (ppm or
mg/L). Values above 1500 ppm (1500 mg/L) are displayed only
in g/L unit. See SETUP section for details.
EC / TDS 27