6.2. CONNECTING THE PROBE

MA815D/1 is connected to the meter through a DIN connector,
making attaching and removing the probe an easy process.
y With the meter o , connect the probe to the DIN socket on the
top of the meter.
y Align the pins and key then push the plug into the socket.
After measurement, power o the meter and clean the probe before
storage.
6.3. ELECTRODE CARE & MAINTENANCE
When using a new probe, remove the sleeve and inspect the
probe prior to use.
Calibrating
Calibration is the first step in obtaining accurate and repeatable
results. See CALIBRATION section for details.
Best practice
y Always use fresh standards. The calibration standards are
easily contaminated.
y Do not reuse standards.
y Do not use expired standards.
Regular Maintenance
y Inspect the probe for cracks or other damage. Replace the
probe if necessary.
y Inspect sensor o-ring for nicks or other damage.
y Inspect the cable. Cable and insulation must be intact.
y Connectors should be clean and dry.
y Follow storage recommendation.
Cleaning Procedure
If a more thorough cleaning is required, remove the sleeve
and clean the probe with a cloth and a nonabrasive detergent.
Reinsert the sleeve and recalibrate the probe.
Storage
EC probes should always be stored clean and dry.
GENERAL OPERATIONS 13