Measurement - Milwaukee MW306 MAX Instruction Manual

Ec / tds / nacl / temperature portable meter
Table of Contents

Advertisement

26 MW

8.3. MEASUREMENT

Conductivity Measurement
When connected, the MA815D/1 probe is automatically recognized.
Place the calibrated probe in the sample, making sure that the
sleeve holes are completely submerged. Tap the probe to remove
any air bubbles that may be trapped inside the sleeve.
To change to EC mode, press RANGE/.
The conductivity value is displayed on the first LCD line,
the temperature on the second LCD line and calibration or
range-specific information on the third LCD line.
To toggle between information displayed on the third LCD line, use
the  keys.
Readings can be temperature compensated.
y Automatic Temperature Compensation (ATC), default: The
probe has a built-in temperature sensor; the temperature value
is used to automatically compensate the EC / TDS reading.
When in ATC mode, ATC tag is displayed and measurements
are compensated using the temperature coe icient.
Recommended default value for water samples is 1.90% / °C.
Temperature compensation is referenced to the selected
reference temperature.