Replacing Control Panel - Invacare Platinum IRC5LXO2AW Service Manual

Hide thumbs Also See for Platinum IRC5LXO2AW:
Table of Contents

Advertisement

Circuit Breaker (STEP 11)
Hour Meter (STEP 11)
1/8" I.D. Pressure
Sensor Tubing (STEP 5)
Oxygen Sensor Tube
(SensO
Models Only)
2
Transformer
Harness
Connector
(STEP 10)
Nine or Ten
Pin
Connector
(STEP 8)
8-Pin Connector

Replacing Control Panel

To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
1. Remove the control panel. Refer to Removing Control Panel on page 50.
2. Remove the P.C. board. Refer to Replacing P.C. Board on page 54.
3. Remove the transformer. Refer to Replacing Transformer Assembly on page 56.
4. Remove the On/Off switch. Refer to Replacing On/Off Switch on page 57.
5. Remove the flow meter. Refer to Replacing Flowmeter on page 58.
6. Remove the hour meter. Refer to Replacing Hour Meter on page 59.
7. Remove horn and palnuts that hold the barbed outlet fitting.
8. Discard the existing control panel.
9. Install new control panel by reversing STEPS 1-7.
10. Run unit and inspect for leaks. Refer to Leak Test on page 71.
11. If no leaks detected, turn power Off (O) and unplug the unit.
12. Reinstall cabinet. Refer to Removing Cabinet on page 25
Part No 1118396
Lower Mounting
(STEP 9)
Screws
FIGURE 13.1 Removing Control Panel
DANGER
SECTION 13—CONTROL PANEL
51
Upper Mounting
Screws
ON/OFF
Switch (I/O)
(STEP 11)
Patient Outlet
Barbed fitting
Top and Bottom
Barbed fittings
Control Panel
®
Platinum
Oxygen