SECTION 20—4-WAY VALVE
11. Remove 4-way valve from manifold assembly. Refer to Replacing 4-Way Valve on
page 60.
12. Remove Y tube and throttle muffler assembly from manifold.
13. Loosen the adjustable clamp on valve manifold input hose to bottom center port of
manifold assembly.
14. Remove valve manifold input hose from bottom center port of manifold assembly.
15. Slide 4-way valve/manifold assembly complete through the cutout in top of sound box
then install the two mounting screws, washer and bushings into the manifold to
secure the 4-way valve assembly to the top of sound box.
16. Reverse STEPS 4-14.
17. Reinstall control panel. Refer to Removing Control Panel on page 50 and reverse all
steps.
18. Run unit and inspect for leaks. Refer to Leak Test on page 71.
19. If no leaks detected, turn power Off (O) and unplug the unit.
20. Reinstall cabinet. Refer to Removing Cabinet on page 25.
®
Platinum
Oxygen
62
Part No 1118396