Konica Minolta CM-36d Instruction Manual page 39

Table of Contents

Advertisement

Symptom
Measurement
²
results are abnormal.
substance in the
(Reflection (color),
transmittance
gloss
²
)
specimen chamber?
Is transmittance
measurement
selected as the
measurement mode?
Measurement
Is the specimen
results are abnormal.
placed properly?
(Transmittance
²
(color))
Is the target mask
installed correctly?
Is there foreign
matter or other dirt
on the inside of the
integrating sphere?
Has correct
calibration data been
written?
Has 0% calibration
been performed
correctly?
Has 100% calibration
been performed
correctly?
Is reflectance
measurement
selected as the
measurement mode?
Data input/output
Is the USB cable
from the instrument
connected correctly?
to the computer is
not possible. No
Is the software
commands from the
operating correctly?
computer are being
accepted.
Check Point
Is there a foreign
²
Remove the specimen or other
substance from the transmittance
specimen chamber.
Refer to the software instruction
manual and perform the
operation correctly.
Set the instrument so that the
sample is in closest possible
contact with the target mask
surface.
Read "Installing the Target
Mask," and install the target mask
correctly.
Foreign matter or dirt larger
than several mm may affect the
measurement results. Foreign
matter and dust inside the
integrating sphere can easily
scratch the barium sulfate paint
on the inside of the sphere. Use
a blower from the specimen
measuring port to blow
these substances off. If such
substances cannot be removed
with the blower, performing zero
calibration and white calibration
may improve performance.
Set the data for the white
calibration plate and gloss
calibration plate
using.
Install the white calibration plate
correctly and perform correct 0%
calibration with light completely
blocked between the illumination
window and receptor window.
Install the white calibration
plate correctly and perform
correct 100% calibration for the
specimen that will be measured.
Refer to the software instruction
manual and perform the
operation correctly.
Connect the instrument and
computer correctly to the USB
cable.
Refer to the software instruction
manual and perform the
operation correctly.
Action
that you are
²
37
Reference Page
30
For details, refer to the
SpectraMagic NX (CM-
S100w) instruction
manual.
30
21
17
For details, refer to the
SpectraMagic NX (CM-
S100w) instruction
manual.
24
24
19
For details, refer to the
SpectraMagic NX (CM-
S100w) instruction manual.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cm-36dg

Table of Contents