Konica Minolta CM-36d Instruction Manual page 37

Table of Contents

Advertisement

Symptom
Displayed message
The specified operation
cannot be executed
because motor drive is in
progress.
There was an operating
failure of the A/D
conversion device inside
the instrument.
There was an operating
failure of light output
circuit charging inside the
instrument.
There was an operating
failure of the light
output circuit inside the
instrument.
There was an operating
failure of a motor inside the
instrument.
There was an operating
failure of memory inside
the instrument.
The camera used to capture
viewfinder images is not
operating.
Possible cause
The operation cannot
be executed because
the motor that performs
switching of the
measurement area and
SCI/SCE is currently
operating.
The A/D conversion device
installed in the instrument
has failed.
Charging of the light
output circuit is not
completed. The light
output circuit is not
operating correctly.
The light output circuit
installed in the instrument
is not operating correctly.
The motor that performs
switching of the
measurement area,
SCI/SCE, and gloss
measurement is not
operating correctly.
The memory installed
in the instrument is not
operating correctly.
The camera installed in the
instrument is not operating
correctly.
35
Correction
Wait for motor operation to
end before executing the
operation.
Turn the power OFF and
then turn it ON again. If
the message is displayed
again, contact a KONICA
MINOLTA-authorized
service facility.
CM-
CM-
36dG
36d