Konica Minolta CM-36d Instruction Manual page 36

Table of Contents

Advertisement

Symptom
Displayed message
Install the white calibration
plate and appropriate
target mask, and then
perform calibration.
Install the gloss calibration
plate and appropriate
target mask, and then
perform calibration.
Light shielding during 0%
calibration was insufficient.
Light intensity at 100%
calibration was insufficient.
A target mask is not
installed. Install the target
mask.
The UV control coefficient
does not exist. Perform
UV control using the UV
Control Software.
Calibration data has not
been set.
Value exceeds the range
that can be measured by
the instrument.
The specified operation
cannot be executed
because measurement is in
progress.
Possible cause
White calibration was
not performed using the
correct procedure.
Gloss calibration was
not performed using the
correct procedure.
0% calibration was not
performed using the
correct procedure.
100% calibration was
not performed using the
correct procedure.
A target mask is not
installed.
Measurement is not
possible because UV
control has not been
performed for the set
conditions.
The calibration data for
white calibration has not
been written.
Color measurement of that
specimen is not possible.
The operation cannot
be executed because
measurement is in
progress.
34
Correction
Perform white calibration
using the provided white
calibration plate and
appropriate target mask.
Perform gloss calibration
using the provided gloss
calibration plate and
appropriate target mask.
Install the white calibration
plate correctly and perform
correct 0% calibration with
light completely blocked
between the illumination
window and receptor
window.
Install the white calibration
plate correctly and perform
correct 100% calibration
for the specimen that will
be measured.
Install the appropriate
target mask before
performing measurement.
Perform UV control for the
set UV conditions from the
UV Control Software that is
provided with SpectraMagic
NX before performing
measurement.
Write the calibration data
to the instrument.
To perform color
measurement, measure
using a different specimen.
Wait for measurement to
end before executing the
operation.
CM-
CM-
36dG
36d