Power management

Helpful hints!
Connect the ventilator to the mains power whenever possible. In the event of battery failure, connect
to mains power immediately to resume ventilation.
An external power source (Astral External Battery or RPSII) should always be available for
ventilator-dependent patients.
An external power source (Astral External Battery or RPSII) should be in use in mobile situations,
including when mains power is unavailable or disrupted. Do not rely solely on the internal battery for
mobile use.
Ensure the external battery is sufficiently charged before using in mobile situations.
WARNING
Beware of electrocution. Do not immerse the device, power supply or power cord in water.
Make sure the power cord and plug are in good condition and the equipment is not damaged.
Keep the power cord away from hot surfaces.
Explosion hazard—do not use in the vicinity of flammable anaesthetics.
The Astral device can be used with different power sources:
Mains power
Astral External Battery
ResMed Power Station II
External DC power supply (eg, car 12V power outlet)
Internal battery
For information on power supplies and sources, refer to the Technical Specifications (see page 73).