Storing and recharging
If the internal battery is not used, it must be recharged every six months.
It takes approximately four hours to fully recharge the internal battery from depletion; however this can
vary depending on environmental conditions and the device operating state.
To prepare the internal battery for long-term storage:
1. Check that the battery charge level is between 50 and 100%. If not, charge the device to at least
50% prior to storage.
2. Remove the power cord from the Astral.
3. Turn off the device.
To recharge the internal battery:
1. Connect the device to mains power.
2. Charging commences as indicated by a flashing battery charging indicator symbol in the Information
bar.
Notes:
When charging a completely depleted battery, it will normally take up to 30 minutes to increase battery capacity
from 0% to 1%.
If the device has been stored outside the operating temperature range, an alarm message (Power fault / No
charging) may appear. You can still continue using the device, however, if the alarm persists for more than 2
hours the battery may need replacement.