Technical specifications
Data storage
Dimensions (L x W x H)
Weight
Inspiratory port / double limb adapter
Pressure measurement
Flow measurement
Power supply
External DC Power Supply
Internal Battery
Housing construction
Environmental conditions
74
7 days of high-resolution airway pressure, respiratory flow and delivered
volume (sampled at 25 Hz).
7 days of breath-related therapy data (sampled at 1 Hz).
365 days of statistical data per program.
285 mm x 215 mm x 93 mm
Display screen size: 150 mm x 90 mm
3.2 kg
22 mm taper, compatible with ISO 5356-1
Anaesthetic & Respiratory Equipment – Conical Connectors
Internally mounted pressure transducers
Internally mounted flow transducers
AC (100-240V), (50-60Hz), 90 W 3.75 A continuous, 120 W / 5A peak
12 - 24V DC 90 W, 7.5 A / 3.75 A
Lithium-Ion battery, 14.4 V, 6.6 Ah, 95 Wh
Operating hours (standard case): 8 hours with a new battery under normal
conditions.
Test conditions: Adult, (A)CV mode, Vt = 800 mL, PEEP = 5 cmH
bpm, Ti = 1.0 sec. All other parameters remain at default settings.
Test lung configuration: R = 5 hPa (L/s)-1, C = 50 mL (hPa)-1
Operating hours (worst case) > 4 hour run time under the following
conditions:
Test conditions: Adult, non-vented, PACV mode, Double limb circuit,
Pressure Assist = 30 cmH
2
Ti: 1.0 sec, Rise Time = Min, Safety Vt = Off, Trig = Off. All other
parameters remain at default settings.
Total lifetime: up to 3,000 hours of normal operation on internal battery
Note: Time may vary with different settings and environmental conditions.
Flame retardant engineering thermoplastic
Operating temperature: 0°C to 40°C
Charging temperature: 5°C to 35°C
Operating humidity: 5 to 93% non-condensing
Storage and transport temperature: -25°C to 70°C for up to 24 hours
Storage and transport temperature: -20°C to 50°C for greater than 24 hours
Note: Storing the Astral device at temperatures exceeding 50°C for
extended period of time may accelerate battery aging. This will not affect
the safety of the battery or device. Refer to Using the internal battery (see
page 46)
Storage and transport humidity: 5 to 93% non-condensing
It takes 40 minutes* for the device to be ready for use on a patient when
removed from storage at the minimum long term temperature and at an
ambient temperature of 20°C.
*Assumes that the device is connected to an external AC power.
It takes 60 minutes for the device to be ready for use on a patient when
removed from storage at the maximum long term temperature and at an
ambient temperature of 20°C.
O, PEEP = 20 cmH
O Rate:20 bpm,
2
O, Rate = 20
2