Agilent Technologies TriScroll 800 Installation And Operation Manual page 41

Inverter dry scroll vacuum pump
Hide thumbs Also See for TriScroll 800:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 49
Table 7
T-Plus Software Variables (Continued)
WIN
TYPE
206
Numerical
211
Numerical
216
Numerical
300
Numerical
301
Numerical
302
Numerical
319
Alphanumeric
323
Alphanumeric
325
Alphanumeric
500
Logical
503
Numerical
504
Logical
Table 8
LED STATUS
Off
Stop
Green flashing
Ramp or Autotuning
Green
Normal Operation
Red
Fail
Orange + Off
Reset + selftest
DESCRIPTION
Error Code:
Bit 7: motor block
Bit 6: short circuit
Bit 5: undervoltage
Bit 4: spare
Bit 3: power fail
Bit 2: controller overt
Bit 1: pump overt
Bit 0: spare
Controller Temperature
[°C]
Environment Temperature
[°C]
Cycle Time
[min]
Cycle Number
Pump Life
[h]
Controller Model
Controller Serial Number
Electrical Modification Level
Monitor Mode (write only)
RS485 Serial Address Setting
[0-31]
Serial Type Select
(0=RS323; 1=RS485)
Status LEDs
CONTROLLER STATUS
NOTE
When the controller is switched on, an
electronic self-test occurs, during
which an orange LED illuminates for
one second and the turns off for two
seconds. During this time, the pump
will not start.
This self-test time is not required if the
pump is already powered, as when
remotely or serially operated.
Using T-Plus Software To Change The Target
Frequency
1.
Install T-Plus software on your PC by running setup.exe.
2.
Using a 9-pin serial cable, connect your PC to the J2
serial port on the inverter (Figure 5).
3.
Start the pump using the On/Off switch. The provided
Interlock Connector must be installed in P1.
4.
Start the T-Plus software and Figure 5 appears.
Figure 5 T+ Pumps List
5.
Click Open Control Panel and Figure 6 appears.
6.
Open the Speed Settings directory, and click Target
Frequency.
Figure 6 T-Unknown Pump
7.
Enter the desired operating frequency into Insert New
Value. This must be an integer between 35 and 65.
8.
Click Send. The pump immediately tries to change to the
new target frequency. If the pump requires more than 750
Watts to attain the target frequency, the pump is power
limited and will run at a lower frequency (auto-tuning
35