Power Cord
Several power cord options are available from your
Agilent dealer. Descriptions of the available power cords and
their ordering numbers are given in Table 1.
NOTE
The pump must be connected to the power
supply using a high voltage IEC-320 type
power cord of at least 10 A capacity.
Table 1 Power Cord Selection
Country
Power Cord Specification
Europe
10 A / 220-230 VAC, 2.5 m
Denmark
10 A / 220-230 VAC, 2.5 m
Switzerland
10 A / 230 VAC, 2.5 m
UK/Ireland
13 A / 230 VAC, 2.5 m
India
10 A / 220-250 VAC, 2.5 m
Israel
10 A / 230 VAC, 2.5 m
North
10 A / 230 VAC, 2.5 m
America
Grounding Instructions
This product should be grounded. In the event of an electrical
short circuit, grounding reduces the risk of electric shock by
providing an escape wire for the electric current. This pump is
equipped with a power cord that has a grounding wire with an
appropriate grounding plug. The plug must be inserted into an
outlet that is properly installed and grounded in accordance
with all local codes and ordinances.
WARNING
Check with a qualified electrician or ser-
viceman if the grounding instructions are
not completely understood, or if you are in
doubt as to whether the product is properly
grounded.
Do not modify the plug provided; if it does
not fit the outlet, have the proper outlet
installed by a qualified electrician.
Connect the product only to an outlet that
has the same configuration as the plug.
Do not use an adapter with this product.
Extension Cords
If you must use an extension cord with this product:
Agilent recommends using only extension cords with a
minimum of 16-gauge wire and a maximum length of 25'
(7.6 m).
Use only a 3-wire extension cord that accepts the plug.
Ensure the extension is in good condition.
Ensure the extension cord is rated high enough to carry the
current the products draws. An undersized cord causes a
drop in line voltage, resulting in loss of power and over-
heating.
Circuit Breakers
Order
WARNING
656494220
656494225
656494235
656494250
656494245
656494230
656494255
Table 2 lists the circuit breaker capacity required for the
pump. Locate the switch or circuit breaker near the pump.
Protect against short circuits by
installing a circuit breaker of the
proper capacity.
The circuit breaker must be suitably
located and easily reached.
The circuit breaker must be marked
at the disconnecting device for the
equipment with the on/off symbol I/O
and meet the relevant requirements of
IEC60947-1 and IEC60947-3.
Table 2
Circuit Breaker Requirements
PTS08001INV
Breaker capacity
10 Amperes
200-240 VAC
50/60 Hz
27