Technical Information - Agilent Technologies TriScroll 800 Installation And Operation Manual

Inverter dry scroll vacuum pump
Hide thumbs Also See for TriScroll 800:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 49

Technical Information

Model
Interface dimensions
Peak pumping speed
Media
Ultimate pressure (Torr)
Maximum inlet pressure
Maximum continuous inlet pressure
Maximum outlet pressure
Inlet connection
Exhaust connection
Gas ballast
Ambient operating temperature
Storage temperature
Motor rating
Operating voltage
Maximum line current
Motor Thermal Protection
Operating speed
Cooling system
Weight
Noise level (per ISO 11201)
Vibration level at inlet (per ISO 10816-1)
Main fuse
Installation and storage
Pollution degree
Altitude
Table 3
Specifications
TriScroll™ 800 Inverter Dry Scroll Vacuum Pump (PTS08001INV)
See Figure 2
3
800 l/m, 48m
/hr (28.3 cfm)
See Figure 3
Clean air. No toxic, corrosive, explosive or particulate forming gases
-3
10.0 x 10
Torr (13.3 x 10
1.0 atmosphere (0 psig)
100 Torr
1.1 atmosphere (1.5 psig)
NW40
Female 3/8" National Pipe Thread (NW25 adapter provided)
Female 1/4" National Pipe Thread (40 Micron sintered filter provided)
5 °C to 40 °C (41 °F to 104 °F)
–20 °C to 60 °C (–4 °F to 140 °F)
1.0 HP (0.75 kW)
50-60 Hz:200-240 VAC Single-phase ±10%, 1200 VA
200 VAC; 6 A
240 VAC; 5A
Automatic
35-65 Hz Factory setting: 62 Hz, 1800 RPM
Air-cooled
Pump only: 32 kg (71 lbs)
Shipping weight: 42 kg (92 lbs)
Variable with frequency, 58-71 dB(A)
Variable with frequency
12.5 A, 250 V
Installation category 2, indoor usage and storage
2
2000 m
-3
mbar)
31

Hide quick links: