Vacuum Products Division TriScroll 300 Series MODULE REPLACEMENT MANUAL Dry Scroll Vacuum Pump Manual No. 699904285 Revision F September 2014...
Page 2
TriScroll™ 300 Dry Scroll Vacuum Pump TriScroll is a trademark of Agilent. Loctite and PST are registered trademarks of Loctite Corporation. Krytox, PTFE and Viton are registered trademarks of E. I. du Pont de Nemours and Company. Copyright 2014 Agilent, Inc.
TriScroll 300 Dry Scroll Vacuum Pump Preface This manual provides the information you need to successfully perform replacement of the entire module of your Agilent TriScroll Dry Vacuum Pump. Module replacement is performed as an alternative to performing a complete overhaul and rebuild (Major Maintenance) and is sold as one of the options recommended when the pump base pressure has risen to an unacceptably high level for your application.
Page 6
TriScroll 300 Dry Scroll Vacuum Pump WARNING Disconnect power from the TriScroll 300 before performing any maintenance procedure. Allow the pump to cool before perform- ing any maintenance procedure. Approximate cool-down time is one to two hours. CAUTION Wipe all O-rings clean with a lint-free cloth before installation to ensure that no foreign matter is present to impair the seal.
TriScroll 300 Dry Scroll Vacuum Pump Related TriScroll Manuals Manuals related to the installation and operation, tip seal replacement, and major maintenance for the TriScroll 300 series pumps are listed in the following table: Title Applicable TriScroll Model Part Number...
Service/Rebuild TriScroll 310 Pump Module Only PTS0310SCRP Serial Numbers Notes This manual applies to TriScroll 300 series with serial numbers beginning with LP, and ascending from LPB80124. For service on TriScroll series pumps with serial numbers J7000001 to A8000108, contact your nearest Agilent office.
TriScroll 300 Dry Scroll Vacuum Pump General Information Agilent TriScroll 300 series pumps are designed to provide years of trouble-free service if maintenance procedures and intervals are observed. Bearing grease replenishment and tip seal replacement is recommended when the pump base pressure has risen to an unacceptably high level for your application.
TriScroll 300 Dry Scroll Vacuum Pump TriScroll Reassembly Tool required: Allen wrench Locate the following items shown in the photos on the left: ➀ Pump Module Assembly ➁ Fan Assembly ➂ M6x16 screws (4) ➃ Spider Coupling ➄ Washer ➅ ➆...
Page 16
TriScroll 300 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 1. Install key in slot on crankshaft. 2. Slide fan assembly onto crankshaft engaging key and against seal spacer. 3. Secure fan assembly to crankshaft with M8x12 screw and washer previously removed in step 5 on page 5.
Page 17
TriScroll 300 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 4. Insert spider in motor coupling. 5. Install pump module on frame, aligning fingers on fan assembly with fingers on coupling. Assure proper fit of dowel pins on mating holes in frame. NOTE Exhaust fitting in the down- ward position.
Page 18
TriScroll 300 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) 6. Secure pump module to frame with four M6x16 screws. 7. Install cowling over pump module. Secure with three M5x16 screws.
TriScroll 300 Dry Scroll Vacuum Pump TriScroll Reassembly (continued) This figure illustrates a fully reassembled TriScroll 300 Series Pump. Pump Conditioning and Performance Verification Pump Module conditioning and performance verification has been performed at Agilent factory. Further conditioning or performance verification is not necessary.
Page 20
TriScroll 300 Dry Scroll Vacuum Pump This page intentionally left blank.
Page 21
Please follow these instructions whenever one of our products needs to be returned. 1) Complete the attached Request for Return form and send it to Agilent Technologies (see below), taking particular care to identify all products that have pumped or been exposed to any toxic or hazardous materials.
Page 22
This page intentionally left blank. 4) HEALTH and SAFETY CERTIFICATION AGILENT TECHNOLOGIES CANNOT ACCEPT ANY PRODUCTS CONTAMINATED WITH BIOLOGICAL OR EXPLOSIVE HAZARDS, RADIOACTIVE MATERIAL, OR MERCURY AT ITS FACILITY. Call Agilent Technologies to discuss alternatives if this requirement presents a problem.
Page 23
Shipping Regulations (IATA, DOT, etc.) and carrier requirements. Customers receiving an Advance Exchange product agree to return the defective, rebuildable part to Agilent Technologies within 15 business days. Failure to do so, or returning a non-rebuildable part (crashed), will result in an invoice for the non-returned/non-rebuildable part.
Page 24
TriScroll 300 Dry Scroll Vacuum Pump This page intentionally left blank.