Agilent Technologies TriScroll 800 Installation And Operation Manual page 40

Inverter dry scroll vacuum pump
Hide thumbs Also See for TriScroll 800:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 49
Serial Interface Specification RS485 and
RS232
Table 5
PIN
#
1
+5V ±5%, 50mA, with reference to pin 5
2
RS232 TX
3
RS232 RX
4
No Connection
5
GND
6
RS485 A
7
No Connection
8
RS485 B
9
No Connection
Physical level: RS232 or RS485
Maximum baud: 9600, 8 data bit, no parity, 1 stop bit.
Table 6
P2 I/O-2 Port on DB9 Male Connector
PIN
#
1
No Connection
2
No Connection
3
No Connection
4
No Connection
5
GND
6
No Connection
7
No Connection
8
No Connection
9
VPI output +24V ±5% 400mA, with reference
to pin 5
34
J1 - Serial Port
SIGNAL NAME
SIGNAL NAME
Table 7
T-Plus Software Variables
WIN
TYPE
000
Logical
001
Logical
008
Logical
102
Numerical
105
Numerical
108
Numerical
117
Numerical
120
Numerical
200
Numerical
201
Numerical
202
Numerical
203
Numerical
205
Numerical
TriScroll 800 Inverter Vacuum Pump
DESCRIPTION
Start/Stop
(1=START; 0=STOP)
Low Speed
[0=OFF / 1=ON]
Remote/Serial Configuration
(1=REMOTE; 0=SERIAL)
Set Point Value
(Hz)
Set Point Hysteresis (%)
[0-100]
Baud Rate
(0-4)
[600-1200-2400-4200-9600]
Low Speed Adjust
(Hz)
Rotational Frequency Setting
[Hz] (High speed adjust)
Bus Current
[mA]
3 Phase Voltage
[Vrms]
Power
[W]
Driving Frequency
[Hz]: (current driving frequency)
Status:
0=stop
2=ramp
3=autotuning
5=normal
6=fail

Hide quick links: