Environmental Conditions, Compliance And Utility Requirements - Agilent Technologies TriScroll 800 Installation And Operation Manual

Inverter dry scroll vacuum pump
Hide thumbs Also See for TriScroll 800:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 49
Environmental Conditions,
Compliance and Utility
Requirements
Environmental Conditions
This equipment meets the following IEC classifications:
Pollution Degree 2 (See Note.)
Installation Category II (See Note.)
Equipment Class 1
NOTE
Pollution level describes the degree to
which a solid, liquid, or gas which deterio-
rates dielectric strength is adhering. 2
applies to a normal indoor atmosphere.
Installation category implies the regula-
tion for impulse withstand voltage. It is
also called the Over voltage category. II
applies to electrical equipment.
This equipment requires the following
space for ventilation, maintenances
access, and easy access to the Main Power
Breaker Switch. There must be a clear
space of at least 60 cm (24") on all sides of
the equipment. The bench in your labora-
tory must be able to support the entire sys-
tem and other laboratory equipment.
Electromagnetic Compatibility
EN55011/CISPR11
Group 1 ISM equipment: Group 1 contains all ISM equip-
ment in which there is intentionally generated and/or used
conductively coupled radio-frequency energy which is neces-
sary for the internal functioning of the equipment itself.
Class A equipment is equipment suitable for use in all estab-
lishments other than domestic and those directly connected to
a low voltage power supply network which supplies buildings
used for domestic purposes.
This device complies with the requirements of CISPR11,
Group 1, Class A as radiation professional equipment. There-
fore, there may be potential difficulties in ensuring electro-
magnetic compatibility in other environments, due to
conducted as well as radiated disturbances.
4
Operation is subject to the following two conditions:
This device may not cause harmful interference.
This device must accept any interference received, includ-
ing interference that may cause undesired operation.
If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the
equipment off and on, the user is encouraged to try one or
more of the following measures:
Make sure that appropriate cables are used to connect the
device to peripheral equipment.
Changes or modifications not expressly approved by Agi-
lent Technologies could void the user's authority to operate
the equipment.
South Korean Class A EMC declaration
This equipment is Class A suitable for professional use and is
for use in electromagnetic environments outside of the home.
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 du Can-
ada.