5.5 Power Requirements
For most installations, the CU 5905 supplies sufficient power for the discussion units. Depending on the setup, the CU
5905 supports anywhere from 22-48 units at a time without additional equipment.
To maximize the unit count without adding equipment:
• Use shorter cables. More units are supported when the distance is minimized between the CU and the discussion
units.
• Use both
outputs from the CU. Units must be divided evenly across each chain.
A B Chain
5.5.1 Units Supported by the CU 5905
The following tables show how to connect the most discussion units without adding additional equipment.
Note: When using two chains, distributable the discussion units evenly across both chains to maximize unit count.
1 Meter between Discussion Units
Distance to DU chain
One Chain
10
44
30
39
50
34
100
21
Figure 19. Power Decreases when Cables Increase
2 Meters between Discussion Units
Two Chains
48
44
40
22
① Length from CU 5905 to first unit
Distance to DU chain
One Chain
10
40
30
36
50
31
100
11
② Distance between participants
Two Chains
48
44
40
22