3
Column Modules
When securing the module from inside the oven do not over tighten. Finger tight is too tight!
C AU T I O N
7 Secure the module with the captive screw(s) from the inside of the LTM oven door. Tighten
Figure 31.
8 Unplug the power supply (or supplies) from mains voltage.
9 Open the electronics enclosure by gently turning the small knob counterclockwise as
10 Route the transfer line and column module connections (total of three cables and
User Guide
enough to establish a reasonable seat against the highly compressible ceramic paper
gasket.
Two column modules installed in LTM oven door (shown after installation of unions)
described in
"Connect the power supply and communications
connections per module) down and through the large slot on the top left of the electronics
housing.
cables".
47