Column Modules
2 If attached, remove the two screws from the ends of transfer line module. These M3 x
Figure 20.
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
WAR N IN G
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
The column module cover mounting screws are very short. Loosen only enough so that the
C AU T I O N
column cover can be removed. Hold the column module over a table or other surface that can
catch a screw that is accidentally removed.
3 Loosen the two screws that secure the column module cover, then tilt to remove the cover.
Column module cover
32
6 mm long screws will be used to attach the two corners of the column module to the ends
of the transfer line module.
Removing the mounting screws