Agilent Technologies LTM II Series User Manual page 46

Rapid heating/cooling system for 8890 gcs
Table of Contents

Advertisement

3
Column Modules
Figure 29.
6 Hold the LTM column module over the fan, then slide forward. Align the clips beneath the
Clip
Tab
Figure 30.
46
The standard module (5-inch) requires two gaskets (one for each slot).
Gaskets installed on a 5-inch module assembly
column module with the tabs on the fan assembly. See
slides towards the LTM oven door, carefully align the unions, brackets, and posts with the
through holes in the LTM oven door. The tolerances are close. Slightly adjust (spread) the
posts if needed.
If you have an assembly that is not clampless, lift up slightly on the module assembly
until the thumbscrew heads of the clamps are cleared, and then the assembly will drop
down into place and slide forward into the oven and the tabs will lock into the slots in the
transfer line module.
Align the module assembly clips over the tabs on the fan bracket
Figure
30. As the column module
User Guide