Agilent Technologies LTM II Series User Manual page 25

Rapid heating/cooling system for 8890 gcs
Table of Contents

Advertisement

3
Column Modules
3 Begin to swage the ferrule to the column using a 1/4-inch open-ended wrench.
Table 4
Ferrule part number
G3188-27501
G3188-27502
G3188-27503
G3188-27504
G3188-27505
G3188-27506
Figure 13.
User Guide
The design of UltiMetal Plus Flexible Metal ferrules reduces the probability of column
breakage and minimizes the damage to the thread of the corresponding fittings.
Unlike other ferrule designs, a reliable leak-free seal is best made by tightening the
internal nut to a recommended number of degrees, not by applying more force. Applying
excessive force will not provide a better seal. Also, compressing the flexible ferrule less
minimizes damage to fittings.
To swage the ferrule to the column:
a Find your ferrule type in
nut.
b Tighten the internal nut clockwise to the lower end of your ferrule's range.
c Check if the ferrule is gripping the column. If yes, stop. If not, continue to tighten the
internal nut in small 5- to 15-degree increments. Check after each increment to see if
the ferrule is gripping the column. Stop as soon as gripping occurs (see
Degrees to tighten the internal nut for proper swaging
Degrees to tighten internal nut
50 – 100 degrees
30 – 70 degrees
20 – 50 degrees
60 degrees
40 – 90 degrees
20 – 50 degrees
Tightening the internal nut
Table 4
and note the range of degrees to tighten the internal
Figure
13).
25