Connections To Electronic Boards; Sensor Cable; Electronic Boards Connectors; Main Board Connections - Hach ORBISPHERE 51 Series User Manual

With k1200 and tc sensors
Hide thumbs Also See for ORBISPHERE 51 Series:
Table of Contents

Advertisement

Installation
3.6

Connections to electronic boards

Note: Any loose connection wires should be bundled tightly together with the use of nylon cable ties.
3.6.1

Sensor cable

An ORBISPHERE cable (10 wire shielded, Part N° 32505.mm) is needed to connect the
sensor(s) to the instrument. The instruments have a Lemo 10 socket on the back panel where
the sensor cable has to be connected.
3.6.2

Electronic boards connectors

Connectors P8 on the main board, and connectors J7 and J8 on the measurement board(s) are
made of two parts. Push down carefully the black levers on either side of the connector and pull
it out securely. Perform all connections with these connectors unplugged. Once finished, attach
the connectors to the boards by pushing them firmly in place (levers up).
3.6.3

Main board connections

Connector P8:
1. RS-485 (signal A)
2. RS-485 (signal B)
3. PROFIBUS-DP (GND)
4. PROFIBUS-DP (+ 5 V)
5. PROFIBUS-DP (signal -)
6. PROFIBUS-DP (signal +)
7. PROFIBUS-DP (signal RTS)
Connector P3
Ethernet RJ 45. Connect the wall and panel mount instruments to the local network by passing
an ethernet cable through the ethernet cable gland (refer to
mount and
above.
30
Potential Instrument Damage. Delicate internal electronic components can be
damaged by static electricity, resulting in degraded performance or eventual failure.
Figure 12 Main board
Figure 10 on page 27
for the panel mount). Connect to the P3 connector illustrated
N O T I C E
Figure 13 Connector P8
8. Not used
9. Not used
10. Not used
11. System alarm relay (N.O.)
12. System alarm relay (N.C.)
13. System alarm relay (Common)
Figure 6 on page 24
for the wall

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orbisphere 511

Table of Contents