External Pressure Calibration; Calibration Reports - Hach ORBISPHERE 51 Series User Manual

With k1200 and tc sensors
Hide thumbs Also See for ORBISPHERE 51 Series:
Table of Contents

Advertisement

Calibration menu
8.4

External pressure calibration

This option is only available if the external pressure option has been enabled for both channels. Refer to
Advanced configuration - K1200 sensor on page 53
Enable channel 1 first, then channel 2.
External pressure calibration
Two calibration methods for the external pressure sensor can be selected:
By default the two point calibration is selected.
Two point calibration
(recommended)
One point calibration
8.5

Calibration reports

Calibration reports
Once a calibration is completed (for a gas or pressure sensor), the calibration report is updated with the
new details. The calibration report contains data for the last 50 calibrations.
For full details on the data displayed for all the different type of calibration reports. Refer to
available on page 77
Each calibration record will contain parameters useful for traceability. For instance, it will contain:
Calibration reports are generated after a sensor calibration. For interfering gas calibration reports, you
will be asked to select the interfering gas from those available.
64
Two point calibration (recommended)
One point calibration
Connect a certified absolute pressure gauge to the sample line, and use a
certified precision barometer.
Note: Any pressure can be used for P1 and P2, but for an accurate
calibration P1 and P2 should be as different as possible.
Connect a certified absolute pressure gauge to the sample line.
Note: Any pressure can be used for P1, but for one point it should be as
close as possible to the sample pressure
for examples.
the user name and ID
the date and time
the calibration coefficient
all the measurements which influence the calibration (temperature, barometric pressure, current,
etc.)
and
Advanced configuration - TC sensor on page
The two point calibration starts with barometric pressure for the lower
point. Expose the external pressure sensor to the atmosphere.
Enter the barometric pressure read on the barometer in the upper
window and validate P1.
Expose the external pressure sensor to line pressure, making sure it is
exposed to the same pressure as the certified absolute pressure gauge.
Enter the absolute pressure value read on the certified absolute
pressure gauge in the lower box, and validate P2.
Expose the external pressure sensor to the line pressure, making sure it
is exposed to the same pressure as the certified absolute pressure
gauge.
Enter the absolute pressure value read on the certified absolute
pressure gauge in lower box, and validate P1.