Tri-linear
Range
1: AOL > I > 4
4-20 mA
2: AOH > I > AOL
3: 20 > I > AOH
1: AOL > I > 0
0-20 mA
2: AOH > I > AOL
3: 20 > I > AOH
1: AOL > U > 0
0-5 V
2: AOH > U > AOL
3: 5 > U > AOH
Analog output characteristics (continued)
None
Measurement M
M=MLL+(ML-MLL)
(I-4)/(AOL-4)
*
M=ML+(MH-ML)
(I-AOL)/ (AOH-AOL)
*
M=MH+(MHH-MH)
(I-AOH) / (20-AOH) R=(MHH-MH)
*
M=MLL+(ML-MLL)
*
M=ML+(MH-ML)
(I-AOL)/ (AOH-AOL)
*
M=MH+(MHH-MH)
(I-AOH)/(20-AOH)
*
M=MLL+(ML-MLL)
U/AOL
*
M=ML+(MH-ML)
(U-AOL)/ (AOH-AOL)
*
M=MH+(MHH-MH)
(U-AOH) / (5-AOH)
*
This is the default value.
Setting the analog output to "None" means that the output value will always
be zero and importantly ensures that no current is emitted on that output, so
reducing power consumption as well as reducing heat within the instrument.
Inputs/Outputs Menu
Resolution R
R=(ML-MLL)
R=(MH-ML)
I/AOL
R=(ML-MLL)
R=(MH-ML)
R=(MHH-MH)
R=(ML-MLL)
R=(MH-ML)
R=(MHH-MH)
20/((AOL-4)
1010)
*
*
20/((AOH-AOL)
1010)
*
*
20/((20-AOH)
1010)
*
*
20/(AOL
1010)
*
*
20/((AOH-AOL)
1010)
*
*
20/((20-AOH)
1010)
*
*
5/(AOL
1010)
*
*
5/((AOH-AOL)
1010)
*
*
5/((5-AOH)
1010)
*
*
73