Installation
3.8.6

Weld-on stainless steel socket

The ORBISPHERE 29501 weld-on sensor socket can be used to install a sensor into a stainless
steel pipe (min.Ø 50 mm or 2"). When welding the socket to the pipe, check that setback
between the pipe's inner diameter and the sensor tip does not exceed 4 mm (see diagram).
1. Pipe external diameter
Note: Be sure to remove the two O-rings from the socket before welding and leave the sensor's stainless
steel cap screwed on during welding to prevent thread distortion.
Recommendation:
To facilitate sensor removal and installation, we suggest installing the socket in a location where
the liquid can be drained quickly and easily. By creating a one meter long piece of pipe
(Figure
enable sensor removal.
Also, a precise sensor and socket installation can be performed in the workshop, and this
assembly can be placed in the production line with minimal down time.
1. Shut off valve
2. Sensor socket
3. Sampling valve
36
Figure 19 Weld-on sensor socket
20) with shut off valves at both ends, just a small volume of liquid needs to be drained to
2. Setback allowed from internal pipe diameter
Figure 20 Installation in-line
4. Shut off valve
5. Sampling valve