Tc Sensor; Tc Sensor Purge Gas Pressure Regulator - Hach ORBISPHERE 51 Series User Manual

With k1200 and tc sensors
Hide thumbs Also See for ORBISPHERE 51 Series:
Table of Contents

Advertisement

Specifications
2.1.3

TC sensor

Sensor type
Membranes
Thickness
Membrane material
Recommended
applications
Radiation limits
Measurement range at 25°C
Accuracy (sample temp. 20-50°C within ± 5°C
of calibration temperature)
Accuracy (sample temp. 0-50°C independent of
calibration temperature)
Measurement cycle (sec.)
Sample flow rate (through flow chamber)
Linear flow rate (past sensor socket)
1
Minimal flow through an ORBISPHERE 32001 flow chamber.
2
Minimal flow past an ORBISPHERE 29501 sensor socket
2.1.4

TC sensor purge gas pressure regulator

14
For other purge gases, please contact your Hach
29561A
25 µm
PFA
Waste gas
offgas,
reactor coolant
10
0-2 ppm, or
0-25 cc/kg, or
0-1.5 bar
The greater of
±1% of reading
or ± 2 ppb,
or ± 0.03cc/kg,
or ± 1.5 mbar
The greater of
±3% of reading
or ± 15 ppb or
± 0.18 cc/kg or
± 6 mbar
1
220 ml/min
2
Model
Filtration grade
Input pressure min/max
Output pressure min/max
Standard nominal flow rate
Condensate volume
Temperature range (ambient and media)
Purge gas dew point
Construction materials
Weight
Size in cm (overall)
31 250 H
and 31260 H
(purge gas N
2
2
representative for specification.
2952A
25 µm
ETFE
Reactor
coolant
5
8
rad
10
rad
0-10 ppm, or
0-120 cc/kg, or
0-6 bar
The greater of
±1% of reading
or ± 8 ppb,
or ± 0.1cc/kg,
or ± 6 mbar
The greater of
±3% of reading
or ± 60 ppb
or ± 0.6 cc/kg,
or ± 20 mbar
17
200 ml/min
N/A
N/A
29089
40 µm
1 bar / 16 bar
0.5 bar / 7 bar
900 Liters/minute
22 cm
-10°C to 60°C
-10°C (+14°F)
Housing: metal
Condensate bowl: polycarbonate
Bowl guard: metal
460 g.
21 x 11.5 x 8
)
31 550 N
2
(purge gas CO
2935A
25 µm
ECTFE
(Halar)
High H
level
2
8
10
rad
0-20 ppm, or
0-350 ppm, or
0-220 cc/kg, or
0-300 ml/l, or
0-12 bar
The greater of
The greater of
±1% of reading
±2% of reading
or ± 25 ppb,
or ± 0.3 ppm,
or ± 0.4 cc/kg,
or ± 0.25ml/l,
or ± 20 mbar
or ± 15 mbar
The greater of
The greater of
±3% of reading
±4% of reading
or ± 150 ppb
or ± 1 ppm
or ± 2.5 cc/kg,
or ± 0.8 ml/l
or ± 50 mbar
or ± 34 mbar
100 ml/min
300 ml/min
N/A
150 cm/sec
3
and
2
31560 N
2
)
2
29561A
25 µm
PFA
In-line
beverage
5
10
rad
0-20 bar
22

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orbisphere 511

Table of Contents