Communication Menu
10.2.2 Input/Output data
The main board:
•
Writes the latest measurement data to the Profibus Input Buffer.
•
Checks if a command written by the Profibus Master must be executed (Profibus Output
Buffer). If a command is to be executed, the instrument executes it and writes the result
(status, data, etc.) in the Profibus Input Buffer.
All numbers are coded in "Big Endian" format, and float values are coded according to IEEE
Standards. The field types "Byte" and "Double Word" are unsigned.
Measurements
Measurements are formatted in the Profibus Input Buffer as follows:
Barometric pressure
Barometric pressure unit
Channel 1 gas concentration
Channel 1 gas unit
Channel 1 temperature
Channel 1 temperature unit
Channel 1 external pressure
Channel 1 external pressure unit
Channel 1 events
Channel 1 measurement index
Channel 2 gas concentration
Channel 2 gas unit
Channel 2 temperature
Channel 2 temperature unit
Channel 2 external pressure
Channel 2 external pressure unit
Channel 2 events
Channel 2 measurement index
Note: The table above shows information for a 2-channel instrument. This will be adjusted according to the
number of channels in use.
82
Name
Type
Input float
Input byte
Input float
Input byte
Input float
Input byte
Input float
Input byte
Input double word
Input double word
Input float
Input byte
Input float
Input byte
Input float
Input byte
Input double word
Input double word
Size
Offset
32 bits
0
8 bits
4
32 bits
5
8 bits
9
32 bits
10
8 bits
14
32 bits
15
8 bits
19
32 bits
20
32 bits
24
32 bits
28
8 bits
32
32 bits
33
8 bits
37
32 bits
38
8 bits
42
32 bits
43
32 bits
47