Installation
3.5

Connection to mains power supply

3.5.1

Power supply connection (low voltage instruments)

For low voltage instruments (10-30 VDC), connection to the mains power supply is with a 7-pin
BINDER connector (supplied).
Connect the power cable to the connector as follows:
3.5.2

Power supply connection (high voltage instruments)

High voltage instruments (100-240 VAC) have a 4-pin male connector pre-wired internally with a
male BINDER connector ready for mains connection. A compatible female connector is supplied
with the instrument.
If this female connector was supplied with a mains power plug already pre-attached (cable part
numbers 33031, 33032, 33033 and 33034) then the female connector can be plugged directly
into the instrument power connector. The two connectors are grooved to avoid an incorrect
fitting. Tighten the female connector to the instrument power connector finger-tight.
If no power cable was ordered with the equipment, a mains power plug must be connected to
the supplied female connector as described in the following procedure.
Personal Injury Hazard. Only qualified personnel should conduct the tasks described in this
section of the manual.
User-supplied power cable specifications:
3-wire (live, neutral and earth)
cable
wire selection mm
Prepare the user-supplied power cable as follows:
28
Figure 11 Wall mount BINDER connector
ø
7mm;  9.5mm
2
, AWG18;  2.5mm
Potential Electrocution Hazard. Always disconnect power to the instrument when
making electrical connections.
Pin Connections:
1. Power 10-30 VDC
2. Ground
3. Ground
4. Ground
5. Not used
6. Power 10-30 VDC
7. Power 10-30 VDC
8. Earth
C A U T I O N
2
, AWG14
WA R N I N G