Section 13 Global Configuration Menu - Hach ORBISPHERE 51 Series User Manual

With k1200 and tc sensors
Hide thumbs Also See for ORBISPHERE 51 Series:
Table of Contents

Advertisement

Section 13 Global Configuration Menu

Global configuration
The global configuration option allows users to save, and use previously saved, instrument
configurations. A maximum of 10 configurations can be saved, with configuration 0 (zero) the instrument
default.
Once all the instrument parameters have been set up, use this option to save the configuration. Selecting
pre-defined configurations avoids the need to re-enter all the parameters when using the instrument for a
different application.
Save
Select
Figure 45 Global configuration menu
Define the index (1-9) to save the current configuration. Use the Next
and Previous buttons to scroll sequentially through existing
configurations, to overwrite an existing configuration or save as a new
one.
Enter a name to define the current configuration.
Enter any comments to associate with this configuration.
Select the configuration (index 0-9) to use on the instrument.
Confirmation will be required for the selected configuration. The
instrument must then be restarted (powered off and then back on) in
order for the new configuration to take effect.
95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orbisphere 511

Table of Contents