Connection To Mains Power Supply; Power Supply Connection (Low Voltage Instruments) - Hach ORBISPHERE G1100 User Manual

Sensor and analyzer
Table of Contents

Advertisement

Installation

3.6 Connection to mains power supply

3.6.1 Power supply connection (low voltage instruments)

For low voltage instruments (10-30 VDC), connection to the mains power supply with a 7-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
30
Figure 12 Wall mount BINDER connector
Pin Connections:
2) Power 10-30 VDC
5) Ground
7) Earth
1+ 3 + 4 + 6) Unused