Installation
3.6 Connection to mains power supply
3.6.1 Power supply connection (low voltage instruments)
For low voltage instruments (10-30 VDC), connection to the mains power supply with a 7-pin
BINDER connector (supplied).
Note: The connectors are grooved to avoid an incorrect fitting to the instrument.
Connect the power cable to the connector as follows:
30
Figure 12 Wall mount BINDER connector
Pin Connections:
2) Power 10-30 VDC
5) Ground
7) Earth
1+ 3 + 4 + 6) Unused