Replacing cover

Advertisement

Available languages

Available languages

Quick Links

Installation Instructions
Promass 80F, 83F, 84F Exd
Replacing cover
EA100D/06/a2/01.10
71106936
Determining the correct spare part set
This method is used merely to determine whether the spare part set is suitable for the measuring device. Other conclusions drawn from the table are not applicable.
1. Read the order code from the nameplate on the measuring device and enter it in the empty fields below.
2. Select the order number of the spare part set in the table below.
3. Check the fields after the order number selected against the highlighted items in the order code.
4. If each field after the selected order number contains a digit or letter from the order code, the spare part set can be used for the measuring device
(* = no restriction).
!
Note! The number of the spare part set can differ from the material number of the spare part it contains!
The order number of the relevant spare part set can be established by entering the material number of the spare part in the spare parts finder.
We recommend that the Installation Instructions be kept with the packaging at all times.
Promass
Order No., Original Spare Part Set
Kit cover 8X with window Inox Exd
71069283
1 × cover D15 complete Inox with screen Exd
1 × O-Ring 123.4 × 3.53 NBR
Kit Cover clamp Inox
2 × claw
71069284
2 × screw M4 × 16
2 × spring washer M4
Kit cover 8X blind version Inox Exd
71069287
1 × cover Inox Exd
1 × O-Ring 123.4 × 3.53 NBR
Kit cover 8X terminal compartment Inox Exd
71069288
1 × cover Inox Exd
1 × O-Ring 113.90 × 3.53 NBR
1 × Kit cover 8X terminal compartment Inox Exe
71069290
1 × cover Inox Exe
1 × O-Ring 113.90 × 3.53 NBR
Determining approval and authorized personnel category:
1. Take the option from the order code field highlighted in gray.
2. Determine the approval and authorized category of personnel using the option in the table.
3. Any person who carries out repairs is responsible for safety during the work, the quality of work completed and safety of the device after repair.
Approval
without Approval
Option
Authorized personnel
1 = Trained specialist on the customer side, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser Flowtec AG (send device)
1* = It is recommended that devices with a design approval be repaired by skilled professionals who have been trained by E+H.
Tool List
24 mm
3 mm
1
8
8
8
8
8
8
ATEX II2G, II1/2G, II2D
"A"
"B, C, D, E, F, G, 1, 2, 3, 4, 5"
1, 2, 3
acid-free,
non-hardening
grease
2
3
4
5
6
7
F * * – * * * * * *
0, 3, 4
F
B, C, D, E, F, K, L, M, G, N, O, P, Q, S, T, U, W, 1, 2, 3, 4, 5, 6
0, 3, 4
F
0, 3, 4
F
B, C, D, E, F, G, K, L, M, N, O, P, Q, S, T, U, W, 1, 2, 3, 4, 5, 6
0, 3, 4
F
B, C, F, K, L, M, N, O, P, U, W, 1, 3, 4
0, 3, 4
F
FM/CSA Div. 1
"N, O, P"
1, 2, 3
8
9
10
11
12
13
*
D, E, G, Q, S, T, 2, 5, 6
TIIS Zone 1
" U, W"
1, 2, 3
1, 2, 3
14
15
16
17
18
*
* * *
L, M, N
L, M, N
L, M, N
L, M, N
L, M, N
NEPSI Zone 1
"K, L, M, S, T, Q"
1, 2, 3

Advertisement

loading

Summary of Contents for Endress+Hauser Promass 80F

  • Page 1 1, 2, 3 1 = Trained specialist on the customer side, 2 = Service technician authorized by Endress+Hauser, 3 = Endress+Hauser Flowtec AG (send device) 1* = It is recommended that devices with a design approval be repaired by skilled professionals who have been trained by E+H.
  • Page 2 – seal the openings with genuine covers – seal cable entries with approved components suitable for the (explosion) protection. • Replace defective seals/gaskets by genuine parts from Endress+Hauser only. • Threads (e.g. of the cover for the electronics and connection compartments) A0011687 must be lubricated.
  • Page 3 1, 2, 3 1 = Ausgebildete Fachkraft des Kunden, 2 = von Endress+Hauser autorisierter Servicetechniker, 3 = Endress+Hauser Flowtec AG (Gerät einsenden) 1* = Es wird empfohlen, Reparaturen welche an Geräten mit Bauartzulassung vorgenommen werden, von ausgebildeten Fachleuten durchführen zu lassen, die von E+H geschult wurden.
  • Page 4 • Um die (Zünd-) Schutzart des Messgeräts wieder herzustellen: – Öffnungen mit Originaldeckeln verschließen – Kabeleinführungen mit zugelassenen, der (Zünd-)Schutzart angepassten Komponenten verschließen. • Defekte Dichtungen nur durch Originaldichtungen von Endress+Hauser ersetzen. • Gewinde (z. B. von Elektronikraum- und Anschlussraumdeckel) müssen A0011687 geschmiert sein.