Endress+Hauser Proline Promass 80E Technical Information
Endress+Hauser Proline Promass 80E Technical Information

Endress+Hauser Proline Promass 80E Technical Information

Coriolis mass flow measuring system mass flow measuring system offering "low cost of ownership" as an alternative to conventional volumetric flowmeters
Hide thumbs Also See for Proline Promass 80E:
Table of Contents

Advertisement

Technical Information

Proline Promass 80E, 83E
Coriolis Mass Flow Measuring System
Mass flow measuring system offering "Low Cost of Ownership" as
an alternative to conventional volumetric flowmeters
Application
The Coriolis measuring principle operates independently
of physical fluid properties, such as viscosity and density.
• Performance characteristics
– Mass flow (liquids):
Promass 80: ±0.30% o.r.
Promass 83: ±0.25% o.r.
– Mass flow (gases):
Promass 80, 83: ±0.75% o.r.
• Process up to +125 °C
• Rupture disc (optional)
• Guaranteed product quality, suitable for CIP/SIP
cleaning
• Low cost of ownership
• Robust field housing (aluminium), IP 67 protection
Approvals in the food industry/hygiene sector:
• 3A authorization
Approvals for:
• ATEX, FM, CSA, TIIS, IECEx, NEPSI
• HART, PROFIBUS DP/PA, FOUNDATION Fieldbus,
MODBUS
• Pressure Equipment Directive, SIL-2
TI061D/06/en/07.07
71044349
Your benefits
The Promass measuring devices make it possible
to simultaneously record several process variables
(mass/density/temperature) for various process
conditions during measuring operation.
The Proline transmitter concept comprises:
• Modular device and operating concept resulting
in a higher degree of efficiency
• Software options for batching and concentration
measurement for extended range of application
• Diagnostic ability and data back-up for increased
process quality
The Promass sensors, tried and tested in over 100000
applications, offer:
• Multivariable flow measurement in compact design
• Insensitivity to vibrations thanks to balanced
two-tube measuring system
• Immune from external piping forces due
to robust design
• Easy installation without taking inlet and outlet
runs into consideration

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass 80E

  • Page 1: Technical Information

    Technical Information Proline Promass 80E, 83E Coriolis Mass Flow Measuring System Mass flow measuring system offering "Low Cost of Ownership" as an alternative to conventional volumetric flowmeters Application Your benefits The Coriolis measuring principle operates independently The Promass measuring devices make it possible of physical fluid properties, such as viscosity and density.
  • Page 2: Table Of Contents

    Proline Promass 80E, 83E Table of contents Function and system design..... 3 Pressure loss ........21 Measuring principle .
  • Page 3: Function And System Design

    Proline Promass 80E, 83E Function and system design Measuring principle The measuring principle is based on the controlled generation of Coriolis forces. These forces are always present when both translational and rotational movements are superimposed. = 2 · Δm (v · ω) = Coriolis force Δm = moving mass...
  • Page 4: Measuring System

    Proline Promass 80E, 83E Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available: • Compact version: transmitter and sensor form a mechanical unit. • Remote version: transmitter and sensor are mounted physically separate from one another.
  • Page 5 Proline Promass 80E, 83E • Single-tube system for highly accurate measurement of Documentation very small flows No. TI 054D/06/en • Nominal diameters DN 1 to 4 • Material: Stainless Steel EN 1.4539/ASTM 904L, EN 1.4404/ASTM 316L (process connection), Alloy C-22 DIN 2.4602 a0003679 •...
  • Page 6: Input

    Proline Promass 80E, 83E Input Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to register a phase shift in the oscillation) • Fluid density (proportional to resonance frequency of the measuring tube) •...
  • Page 7: Output

    Proline Promass 80E, 83E Output Output signal Promass 80 Current output: Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA < 700 Ω (for HART: R ≥...
  • Page 8 Proline Promass 80E, 83E PROFIBUS PA interface: • PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated • Data transmission rate: 31.25 kBit/s • Current consumption: 11 mA • Permitted supply voltage: 9 to 32 V •...
  • Page 9: Signal On Alarm

    Proline Promass 80E, 83E Signal on alarm Current output: Failsafe mode selectable (e.g. in accordance with NAMUR Recommendation NE 43) Pulse/frequency output: Failsafe mode selectable Status output (Promass 80) Nonconductive in the event of a fault or if the power supply fails...
  • Page 10: Power Supply

    Proline Promass 80E, 83E Power supply Electrical connection Measuring unit b d/(g) HART* PROFIBUS PA* FOUNDATION Fieldbus* – 27 PA(–)/FF(–) + 26 PA(+)/FF(+) – – 25 + 24 – – 23 + 22 – – 21 + 20 N (L-)
  • Page 11: Electrical Connection, Terminal Assignment

    Proline Promass 80E, 83E Electrical connection, Promass 80 terminal assignment Terminal No. (inputs/outputs) Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–) 80***-***********A Frequency output Current output, HART...
  • Page 12: Electrical Connection Remote Version

    Proline Promass 80E, 83E Terminal No. (inputs/outputs) Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–) Current output 83***-***********L Status input Relay output 2 Relay output 1...
  • Page 13: Cable Entries

    Proline Promass 80E, 83E Cable entries Power-supply and signal cables (inputs/outputs): • Cable entry M20 × 1.5 (8 to 12 mm) • Thread for cable entries, ½" NPT, G ½" Connecting cable for remote version: • Cable entry M20 × 1.5 (8 to 12 mm) •...
  • Page 14 Proline Promass 80E, 83E Promass 83 ±0.25% ± [(zero point stability ÷ measured value) · 100]% o.r. Mass flow (gas) ±0.75% ± [(zero point stability ÷ measured value) · 100]% o.r. Volume flow (liquid) ±0.45% ± [(zero point stability ÷ measured value) · 100]% o.r.
  • Page 15: Repeatability

    Proline Promass 80E, 83E Repeatability Mass flow (liquid) ±0.10% ± [½ · (zero point stability ÷ measured value) · 100]% o.r. Mass flow (gas) ±0.35% ± [½ · (zero point stability ÷ measured value) · 100]% o.r. Volume flow (liquid) ±0.20% ±...
  • Page 16: Operating Conditions: Installation

    Proline Promass 80E, 83E Operating conditions: Installation Installation instructions Note the following points: • No special measures such as supports are necessary. External forces are absorbed by the construction of the instrument, for example the secondary containment. • The high oscillation frequency of the measuring tubes ensures that the correct operation of the measuring system is not influenced by pipe vibrations.
  • Page 17 Proline Promass 80E, 83E Orientation Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow (direction of fluid flow through the pipe). Vertical (view V) Recommended orientation with upward direction of flow. When fluid is not flowing, entrained solids will sink down and gases will rise away from the measuring tube.
  • Page 18 Proline Promass 80E, 83E Special installation instructions for Promass P " Caution! When using a bent measuring tube and horizontal installation, the position of the sensor has to be matched to the fluid properties! a0004581 Horizontal installation for sensors with a bent measuring tube Not suitable for fluids with entrained solids.
  • Page 19: Inlet And Outlet Runs

    – Relative magnetic permeability μ – Plate thickness d ≥ 0.35 mm • Information on permitted temperature ranges → Page 20 Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available for the sensors. Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs.
  • Page 20: Operating Conditions: Environment

    Proline Promass 80E, 83E Operating conditions: Environment Ambient temperature range Standard: –20 to +60 °C (sensor, transmitter) Optional: –40 to +60 °C (sensor, transmitter) Note! • Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
  • Page 21: Pressure Loss

    Proline Promass 80E, 83E Pressure loss Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to approximately calculate the pressure loss: 2 · Re = Reynolds number · d · ·...
  • Page 22: Mechanical Construction

    Proline Promass 80E, 83E Mechanical construction Design / dimensions Dimensions: → Page 23 Field housing compact version, powder-coated die-cast aluminum → Page 24 Dimensions: Remote version → Page 24 Transmitter connection housing remote version (II2G/Zone 1) → Page 25 Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2) →...
  • Page 23 Proline Promass 80E, 83E Field housing compact version, powder-coated die-cast aluminum A0007638 All dimensions in [mm]; * Blind version (without local display) All dimensions in [mm]; * dependent on respective process connection → For dimensions, see the following pages Note! Dimensions for transmitters II2G/Zone 1 →...
  • Page 24 Proline Promass 80E, 83E Dimensions: Remote version a0007637 All dimensions in [mm] Transmitter connection housing remote version (II2G/Zone 1) Nicht-eigensichere Stromkreise durch IP40-Abdeckung geschützt Non-intrinsically safe circuits Ip40 protected Boucles de courant sans sécurité intrinsèque protégées par Ip40 a0002128 * Blind version (without local display) Ø...
  • Page 25 Proline Promass 80E, 83E Transmitter wall-mount housing (non Ex-zone and II3G/Zone 2) a0001150 90.5 159.5 >50 81.5 11.5 8xM5 All dimensions in [mm] Endress+Hauser...
  • Page 26 Proline Promass 80E, 83E Flange connections EN (DIN), ASME B16.5, JIS +1.5 –2.0 a0007640-en Flange EN 1092-1 (DIN 2501 / DIN 2512N / PN 40: 1.4404/316L 4 × Ø14 17.3 5.35 4 × Ø14 17.3 8.30 4 × Ø14 28.5 12.00...
  • Page 27 Proline Promass 80E, 83E Flange according to ASME B16.5 / Cl 150: 1.4404/316L 3/8" 88.9 4 × Ø15.7 11.2 60.5 15.7 5.35 ½" 88.9 4 × Ø15.7 11.2 60.5 15.7 8.30 1" 108.0 4 × Ø15.7 14.2 79.2 26.7 12.00 1½"...
  • Page 28 Proline Promass 80E, 83E Flange JIS B2220 / 40K: SUS 316L 4 × Ø19 5.35 4 × Ø19 8.30 4 × Ø19 12.00 4 × Ø23 17.60 8 × Ø19 26.00 All dimensions in [mm]; Further dimensions → Page 23 ff.
  • Page 29 Proline Promass 80E, 83E Tri-Clamp connections +1.5 –2.0 a0007643-en 1", 1½", 2" Tri-Clamp: 1.4404/316L Clamp 1" 50.4 22.1 5.35 1" 50.4 22.1 8.30 1" 50.4 22.1 12.00 1½" 50.4 34.8 17.60 2" 63.9 47.5 26.00 All dimensions in [mm]; Further dimensions → Page 23 ff.
  • Page 30 Proline Promass 80E, 83E DIN 11851 (hygienic connection) +1.5 –2.0 a0007644-en Hygienic connection DIN 11851: 1.4404/316L Rd 34 × 1/8" 5.35 Rd 34 × 1/8" 8.30 Rd 52 × 1/6" 12.00 Rd 65 × 1/6" 17.60 Rd 78 × 1/6"...
  • Page 31 Proline Promass 80E, 83E DIN 11864-2 Form A (flat flange) +1.5 –2.0 a0007649-en flat flange Flange DIN 11864-2 Form A ( ): 1.4404/316L 4 × Ø9 5.35 4 × Ø9 8.30 4 × Ø9 12.00 4 × Ø9 17.60 4 × Ø9 26.00...
  • Page 32: Rupture Disk In The Sensor Housing (Optional)

    Proline Promass 80E, 83E SMS 1145 (hygienic connection) +1.5 –2.0 a0007653-en Hygienic connection SMS 1145: 1.4404/316L Rd 40 × 1/6" 22.5 5.35 Rd 40 × 1/6" 22.5 8.30 Rd 40 × 1/6" 22.5 12.00 Rd 60 × 1/6" 35.5 17.60 Rd 70 ×...
  • Page 33: Weight

    Proline Promass 80E, 83E Weight • Compact version: see table below • Remote version – Sensor: see table below – Wall-mount housing: 5 kg Compact version Remote version (sensor) All values (weight) refer to devices with EN/DIN PN 40 flanges.
  • Page 34: Material Load Curves

    Proline Promass 80E, 83E Material load curves Flange connection to EN 1092-1 (DIN 2501) Flange material: 1.4404/316L [bar] PN 100 PN 63 PN 40 [°C] -60 -40 -20 80 100 120 140 180 200 a0006904-en Flange connection according to ASME B16.5 Flange material: 1.4404/316L...
  • Page 35 Proline Promass 80E, 83E VCO process connection Flange material: 1.4404/316L [bar] PN 100 -60 -40 -20 80 100 120 140 180 200 [°C] a0006908-en Tri-Clamp process connection The Clamp connections (e.g. Tri-Clamp ISO2852, DIN32676) are suited up to a maximum pressure of 16 bar.
  • Page 36: Process Connections

    Proline Promass 80E, 83E Coupling to ISO 2853 Coupling material: 1.4404/316L [bar] -60 -40 -20 80 100 120 140 180 200 [°C] A0006912-en See Page 33 → Materials → Process connections Process connections Human interface Display elements • Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line •...
  • Page 37: Certificates And Approvals

    CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-Tick symbol The measuring system complies with the EMC requirements of the "Australian Communications and Media Authority (ACMA)"...
  • Page 38: Functional Safety

    Promass83***-************6 Ordering information The Endress +Hauser service organization can provide detailed ordering information and information on the order codes upon request. Accessories Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor. Endress+Hauser...
  • Page 39: Documentation

    ® MODBUS Registered trademark of the MODBUS Organization ® ® ® ® ® HistoROM™, S-DAT , T-DAT™, F-CHIP , ToF Tool - Fieldtool Package, Fieldcheck , Applicator ® FieldCare Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH Endress+Hauser...
  • Page 40 Instruments International Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel. +41 61 715 81 00 Fax +41 61 715 25 00 www.endress.com info@ii.endress.com TI061D/06/en/07.07 71044349 FM+SGML6.0 ProMoDo...

This manual is also suitable for:

Proline promass 83e

Table of Contents