Bosch BEA 150 Repair Instructions page 19

Emissions analysis
Table of Contents

Advertisement

4.7.5
Readjustment with test gas
i A distinguishing feature of the exhau-
st-gas analyzer module is the excel-
lent long-time stability of its measu-
ring accuracy. Despite this, the ex-
haust-gas analyzer may be required
by law or by the regulations of appro-
val authorities to be recalibrated with
calibrating gas at given intervals.
To recalibrate the exhaust-gas analyzer,
you will need a calibrating-gas mixture
with the following concentrations (depen-
ding on requirements):
HC:
400 to 2000 ppm vol C
entry in propane
CO:
1% to 10% vol CO
CO
: 5% to 18% vol CO
2
Using the keyboard, enter the calibrating-
gas values for CO as stated on the certi-
ficate.
Confirm your entry of these values by
clicking OK or by pressing E.
Using the keyboard, enter the calibrating-
gas values for CO
as stated on the certifi-
2
cate.
Confirm your entry of these values by
clicking OK or by pressing E.
Using the keyboard, enter the calibrating-
gas values for HC as stated on the certif-
icate.
Confirm your entry of these values by
clicking OK or by pressing E.
H
,
3
8
2
19