Sensor Available; Sensor Not Available - Bosch BEA 150 Repair Instructions

Emissions analysis
Table of Contents

Advertisement

4.5.6
O

sensor available

2
4.5.7
O

sensor not available

2
4.5.8
Entering lambda coefficients
OC
and HC
V
V
16
Open the O
sensor available menu using
2
the oukeys and confirm with E or by
double-clicking the mouse.
After confirmation that the O
available, the measuring chamber's
measuring computer (SI BENCH) is re-
started. After this it jumps back to the Set
parameters sub menu.
i The exhaust-gas analyzer is then once
more in the warm-up phase. Further
measurement with the Service soft-
ware is disabled for 60 seconds. A
message appears on screen.
Open the O
sensor not available menu
2
using the ou keys and confirm with E or
by double-clicking the mouse.
After confirmation that the O
available, the measuring chamber's
measuring computer (SI BENCH) is re-
started. After this it jumps back to the Set
parameters sub menu.
i The exhaust-gas analyzer is then once
more in the warm-up phase. Further
measurement with the Service soft-
ware is disabled for 60 seconds. A
message appears on screen.
i Lambda coefficients must also be set
in accordance with regulations speci-
fic to the country of use and must be
stipulated by the law of that country.
Lambda coefficients OC
are fuel-specific values and are re-
quired to calculate the lambda value.
Enter the values for OC
on the keyboard
V
and confirm by clicking OK or by press-
ing E.
e.g.: 0.0175.
Enter the values for HC
on the keyboard
V
and confirm by clicking OK or by press-
ing E.
e.g.: 1.7261.
sensor is not
2
sensor is not
2
and HC
V
V