The digitizer is connected to the workstation via Ethernet connection and uses
the DICOM protocol to communicate with the workstation.

Installation

The digitizer is equipped with 2 handles at the bottom left and right sides to
move the device easily to another location. It is recommended to have at least
two persons lift the digitizer.
16
Introduction
Caution:
When installing the digitizer, care must be taken to ensure that
there is either a mains plug or an all-cable disconnecting device in the
internal installation fitted near the digitizer and that it is easily
accessible.
Caution:
The digitizer and the cassette storage shall be protected against
direct radiation in such a way, that the annual dose equivalent at the
place of installation will not exceed 1 mSv/a.
Caution:
Do not lift the device by holding the input tray.
Caution:
If the digitizer is installed inside an X-ray room it must be
protected from stray radiation by proper shielding.
2386G EN 20120110