Questionable Calibration Register

Questionable Calibration Condition
:STATus:QUEStionable:CALibration:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Calibration Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
Key Type:
Questionable Calibration Enable
:STATus:QUEStionable:CALibration:ENABle <number>
:STATus:QUEStionable:CALibration:ENABle?
This command determines which bits in the Questionable Calibration
Condition Register will set bits in the Questionable Calibration Event
register, which also sets the Calibration Summary bit (bit 8) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Key Type:
Example
Factory Preset: 32767 (all 1's)
Range:
Chapter 6
There is no equivalent front-panel key.
0 to 32767
There is no equivalent front-panel key.
There is no equivalent front-panel key.
STAT:QUES:CAL:ENABLE 16384 could be used if you
have turned off the automatic alignment and you want
to query if an alignment is needed.
0 to 32767
Using the STATus System
STATus Subsystem