Questionable Calibration Condition

:STATus:QUEStionable:CALibration:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Calibration Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.

Questionable Calibration Enable

:STATus:QUEStionable:CALibration:ENABle <number>
:STATus:QUEStionable:CALibration:ENABle?
This command determines what bits in the Questionable Calibration
Condition Register will set bits in the Questionable Calibration Event
register, which also sets the Calibration Summary bit (bit 8) in the
Questionable Register. The variable <number> is the sum of the
decimal values of the bits you want to enable.
Factory Preset
and *RST:
Range:

Questionable Calibration Event Query

:STATus:QUEStionable:CALibration[:EVENt]?
This query returns the decimal value of the sum of the bits in the
Questionable Calibration Event register.
The register requires that the equivalent PTR or NTR filters be set
NOTE
before a condition register bit can set a bit in the event register.
The data in this register is latched until it is queried. Once queried, the
data is cleared.
Chapter 5
32767 (all 1's), factory default. (The user setting is
persistant.)
0 to 32767
Language Reference
STATus Subsystem
269