Questionable Status Register - Agilent Technologies 4155C Command Reference Manual

Semiconductor parameter analyzer precision semiconductor parameter analyzer
Hide thumbs Also See for 4155C:
Table of Contents

Advertisement

Figure 3-7
Status Reporting Structure

Questionable Status Register

Questionable Status Register
This register consists of CONDITION, TRANSITION FILTER, EVENT, and event
enable (MASK) registers.
You enable the desired bits of the EVENT register by using the MASK register.
1 enables, 0 masks the corresponding bit of EVENT register. The MASK
register is set by the :STATus:QUEStionable:ENABle command.
The TRANSITION FILTER register consists of positive and negative transition
filters.
If a bit in the positive transition filter is set to 1, then a 0 to 1 transition in the
corresponding bit of the CONDITION register causes a 1 to be written in the
corresponding EVENT register bit. The positive transition filter is set by the
:STATus:QUEStionable:PTRansition command.
If a bit in the negative transition filter is set to 1, then a 1 to 0 transition in the
corresponding bit of the CONDITION register causes a 1 to be written in the
corresponding EVENT register bit. The negative transition filter is set by the
:STATus:QUEStionable:NTRansition command.
See "STATus Subsystem" in Chapter 5 for details about these commands.
Questionable Status Register
This is a standard SCPI register, but non-IEEE488.2.
3-12
Agilent 4155C/4156C SCPI Command Reference, Edition 1

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

4156c

Table of Contents