Agilent Technologies PSA Series User And Programming Manual page 326

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Programming Fundamentals
Using the LAN to Control the Instrument
Common Problems After a Connection is Made
This section describes common problems you may encounter when
using the instrument on a LAN. It assumes you have been able to
connect to the instrument in the past. If this is not so, refer to the
previous sections first.
Pressing
NOTE
Restore Sys Defaults
then have to re-set the instrument IP address and other LAN settings
in
System, Config I/O
Remember that in any type of programming using LAN you should
NOTE
avoid constantly opening and closing connections. This uses up
processing resources, adds to your system overhead, and can cause
problems with asynchronous implementation of successive commands.
When you are sending the instrument multiple commands: open the
connection, send all the commands, and close the connection.
Cannot connect to the analyzer
• If you suspect a bad LAN connection between your computer and
• If a bad connection is revealed, try the following solutions:
326
does not affect LAN settings, but pressing
Preset
will reset to the original factory defaults. You will
.
instrument, you can verify the network connection by using the ping
command described later in this chapter or another similar echo
request utility.
— Make sure the instrument is turned on.
— Check the physical connection to the LAN.
— Make sure the internet (IP) Address of the instrument is set up
correctly in the LAN port setup menu. (Press
.)
Address
— When connecting to your instrument over a closed network
(directly through a hub or crossover cable) it may help to set the
instrument to its default settings for subnet mask and gateway.
(subnet mask: 255.255.0.0, gateway 0.0.0.0)
— If the instrument and the computer are on different networks or
subnets, make sure the gateway address and subnet mask values
are set correctly.
System,
System, Config I/O, IP
Chapter 5

Advertisement

Table of Contents
loading

This manual is also suitable for:

E4440aE4443aE4445aE4448aE4446a

Table of Contents