Power Meter Remote Operation
Table 1-7 Bit Definitions - Questionable Status Registers
The condition bits are set and cleared under the following conditions:
Table 1-8 Bit change conditions for Questionable Status Register
62
Bit
Decimal
Number
Weight
0 to 2
-
3
8
4 to 7
-
8
256
9
512
10 to 14
-
15
-
Bit
Meaning
Number
3
POWer
Summary
Definition
Not used
POWer Summary
Not used
CALibration Summary
Power On Self Test
Not Used
Not used (always 0)
EVENts Causing Bit Changes
This is a summary bit for the Questionable POWer Register.
•
SET:
Error –230, "Data corrupt or stale"
Error –231, "Data questionable;Input Overload"
Error –231, "Data questionable;Input Overload ChA"
Error –231, "Data questionable;Input Overload ChB"
Error –231, "Data questionable;PLEASE ZERO"
Error –231, "Data questionable;PLEASE ZERO ChA"
Error –231, "Data questionable;PLEASE ZERO ChB"
Error –231, "Data questionable;Lower window log error"
Error –231, "Data questionable;Upper window log error"
•
CLEARED: When no errors are detected by the power
meter during a measurement covering the causes given
for it to set.
N1911A/1912A P-Series Power Meters Programming Guide
*
*
*
*
*
*