Status Registers
Use Status Registers to Determine the State of Analyzer Events and Condition s
Bit
Decimal
Value
Description
4
16
synth unlocked A 1 in this bit position indicate s
that the analyzer synthesizer is unlocked .
5
32
Invalid BW A 1 in this bit position indicates that an
invalid bandwidth setting has been requested .
6-8
-
Reserved These bits are not used by the analyzer, but
are for future use with other Agilent products.
9-14
-
Unused These bits are always set to 0 .
15
32768
Always zero (0) This bit is always set to 0.
STATus:QUEStionable:CALibration Condition
Register Bits
Bit
Decima l
Value
Description
0, 1
0, 2
Reserved These bits are not used by the analyzer, but
are for future use with other Agilent products .
2
4
TG Align Failure A 1 in this bit position indicate s
that a failure has occurred while trying to align th e
tracking generator (TG) .
3
8
RF Align Failure A 1 in this bit position indicate s
that a failure has occurred while trying to align th e
RF section .
4
16
IF Align Failure A 1 in this bit position indicate s
that a failure has occurred while trying to align the I F
section.
5
32
LO Align Failure A 1 in this bit position indicate s
that a failure has occurred while trying to align th e
local oscillator (LO) .
6
64
ADC Align Failure A 1 in this bit position indicate s
that a failure has occurred while trying to align th e
analog-to-digital converter (ADC) .
7
128
FM Demod Align Failure A 1 in this bit position
indicates that a failure has occurred while trying t o
align the FM demodulation circuitry. (Agilent ESA
models E4401B, E4402B, E4404B, E4405B, an d
E4407B Option BAA, FM Demodulation, only) .