*PSC, *PSC? (power-on state clear) sets the power-on state so that it
clears the service request enable register and the event status
enable register at power on.
*SRE, *SRE? (service request enable) sets and queries the value of
the service request enable register.
*STB? (status byte) queries the value of the status byte register
without erasing its contents.

Why Use the Status Registers

A program often needs to be able to detect and manage error conditions
or changes in instrument status. There are two methods you can use to
programmatically access the information in status registers:
• The polling method
• The service request (SRQ) method
In the polling method, the instrument has a passive role. It only tells
the controller that conditions have changed when the controller asks
the right question. In the SRQ method, the instrument takes a more
active role. It tells the controller when there has been a condition
change without the controller asking. Either method allows you to
monitor one or more conditions.
The polling method works well if you do not need to know about
changes the moment they occur. The SRQ method should be used if you
must know immediately when a condition changes. To detect a change
using the polling method, the program must repeatedly read the
registers.
Use the SRQ method when:
— you need time-critical notification of changes
— you are monitoring more than one device which supports SRQs
— you need to have the controller do something else while waiting
— you can't afford the performance penalty inherent to polling
Use polling when:
— your programming language/development environment does not
support SRQ interrupts
— you want to write a simple, single-purpose program and don't want
the added complexity of setting up an SRQ handler