Supplemental Electrical Characteristics; Residual Errors After Calibration; Certification - Agilent Technologies 85056K Manual

Table of Contents

Advertisement

Specifications
Electrical Specifications
b.
Assuming proper usage, the residual return loss after calibration specifications for
the sliding load termination includes the quality of the airline portions within the
sliding load, combined with the effective stability of the sliding element. Proper usage
includes the following practices: (1) Connector mating surfaces are clean; (2) The
changes in slide positioning are NOT done in equal steps — equal steps results in very
poor calibration for some portions of the frequency range; (3) The center conductors of
testport connectors are nominally set back from the outer conductors. Sliding loads are
designed to allow the center conductor to be moved. The position of the sliding load
center conductor should be set by a reference block and not positioned flush against
the center conductor of the testport.
c.
The 2.4 mm to 2.92 mm adapters are tested two at at time (connected together) at the factory.
d. The specifications for the opens and shorts are given an allowed deviation from the nominal
model as defined in the standard definitions (see
Definitions Values are Available on the Web" on page A-2

Supplemental Electrical Characteristics

Table 2-4 on page 2-6
adapters in this kit. Values in this table are not specifications, but are intended to provide
useful application information by giving typical, but non-warranted, performance
parameters.
Table 2-4
2.4 mm to 2.92 mm adapter Characteristics
Frequency (GHz)
DC to
2
2 to
20
>
20 to
40
>
DC to
40
DC to
40

Residual Errors after Calibration

The 8510 "Specifications and Performance Verification" software can be used to obtain a
printout of the residual errors after a calibration has been performed. Refer to the
"Specifications and Performance Verification" section of the 8510 On-Site Service Manual
for information on how to use the software.

Certification

Agilent Technologies certifies that this product met its published specifications at the time
of shipment from the factory. Agilent further certifies that its calibration measurements
are traceable to the United States National Institute of Standards and Technology (NIST)
to the extent allowed by the institute's calibration facility, and to the calibration facilities
of other International Standards Organization members. See
Devices in Your Kit," on page 4-2
2-6
"Class Assignments and Standard
lists the typical electrical characteristics of the 2.4 mm to 2.92 mm
Parameter
Typical Value
Return Loss
Return Loss
Return Loss
Electrical Length
39.631 ps
Insertion Loss
<
for more information.
).
0.01259 ρ)
38 dB (
0.01778 ρ)
35 dB (
0.03162 ρ)
30 dB (
±
0.14 ps
0.99140 ρ)
0.075 dB (
>
"How Agilent Verifies the
85056K