Software Setup

4.4 Software Setup for Titrant Installation

1.
From the Titrants menu,
press
to select Install
2
Titrants.
5.
Highlight Expiration.
Press the
to
CHECK MARK
input or change.
Software Setup
Page 38
2.
Press
to select
1
Titrants Setup.
6.
Use the alphanumeric
keypad to input the month
and year of the titrant
expiration date (located next
to the barcode on the titrant
bottle).
Press the
to
CHECK MARK
validate.
3.
Highlight Lot no.
Press the
to
CHECK MARK
input or change.
7.
Verify the titrant
information. Press
to
1
confirm.
4.
Use the alphanumeric
keypad to input or change
the titrant lot number—
located next to the barcode
on the titrant bottle.
Press the
to
CHECK MARK
validate.
8.
Press
to select
2
No Rinse.
Note: Initial installation does
not require a burette rinse;
however, it is highly
recommended for titrant
replacement or exchange.
50081_SoftwareSetup.fm