Hach AutoCAT 9000 Instruction Manual page 72

Chlorine amperometric titrator
Table of Contents

Advertisement

Chlorine Dioxide Generator Yield
13.
Lower the electrode
head onto the beaker rim.
Press the
CHECK MARK
to confirm.
Chlorine Dioxide Generator Yield
Page 70
14.
The analysis parameters
that have been stored for the
current Sub-ID may be
edited at this time.
Highlight the appropriate
line and press the
to change one
CHECK MARK
of the analysis parameters or
press 1 to confirm.
Note: Any changes made will
become the new default
parameters for this method
Sub-ID.
Note: These Sub-ID
parameters can be edited here
as 'run-time' settings, or from
the Main Menu as described in
Section 3.5.
Note: This step will be skipped
if auto-scaling software is
enabled.
15.
The Start Timer
will start to allow the
electrode to stabilize before
data collection.
Once the electrode has
stabilized, the AutoCAT will
begin data acquisition and
plot the titration curve.
16.
The Zoom Range
Determination window will
display the titration curve.
Use the navigation buttons to
select and move the cursors
to define the range of data
points for analysis.
Note: Defining the range of
data points excludes
undesirable portions of the
titration curve from
the analysis.
50081_ChlorineDioxideGeneratorYield.fm