Hach AutoCAT 9000 Instruction Manual page 106

Chlorine amperometric titrator
Table of Contents

Advertisement

Total Chlorine
Summary of Method
Required Reagents
Description
Acetate Buffer Solution, pH 4, w/dropper................................................... 1 mL................. 100 mL ...............14909-32
Iodine Standard Solution, 0.0282 N ............................................................ varies................ 1000 mL .............23333-53
Phenylarsine Oxide (PAO) Titrant, 0.00564 N............................................ varies................ 1 L ........................1999-53
Potassium Iodide, SwifTest refill ................................................................ 0.10 g ............... 25 g.......................1077-60
Required Apparatus
Beaker, Glass, 250-mL ................................................................................ 1 ....................... each......................500-46H
Cylinder, Graduated, 250 mL ...................................................................... 1 ....................... each.........................508-46
Magnetic Stir Bar, Teflon
Pipet, Volumetric, Class A, 1.00 mL ........................................................... 1 ....................... each.....................14515-35
SwifTest Dispenser...................................................................................... 1 ....................... each.....................28341-00
Optional Reagents
Chlorine Standard Solution, Voluette
Dilution Water, organic-free ........................................................................ varies................ 500 mL ...............26415-49
Phenylarsine Oxide (PAO) Titrant, 0.00564 N, cartridge ........................... varies................ each.......................1949-01
Sodium Thiosulfate Standard Solution, stabilized, 0.00564 N ................... varies................ 1 L ......................24088-53
Optional Apparatus
Digital Titrator....................................................................................................................... each.....................16900-01
Delivery Tubes, straight w/"J" hook ..................................................................................... 5/pkg...................17205-00
Mini-printer, thermal, w/cable............................................................................................... each.....................A70P021
*Teflon is a Registered Trademark of EI DuPont Company.
FOR TECHNICAL ASSISTANCE, PRICE INFORMATION AND ORDERING:
In the U.S.A. –
Outside the U.S.A. –
On the Worldwide Web –
© Hach Company, 2002–2003, 2007. All rights reserved. Printed in the U.S.A.
Page 104
This procedure follows method number 4500-Cl c. in "Standard Methods for the
Examination of Water and Wastewater" (20th Edition). In the amperometric back
titration, a measured volume of reductant (phenylarsine oxide or sodium thiosulfate
solution) is added directly to a measured volume of the sample at collection time. This
"fixes" the chlorine concentration, allowing the sample to be stored for later analysis.
At analysis time, the sample is adjusted to pH 4 with acetate buffer, and excess potassium
iodide is added. The unreacted reductant is then titrated with standard tri-iodide solution.
The chlorine concentration is derived from the difference in the amount of reductant
originally added to the sample and the amount remaining prior to titration.
The chemical reactions are:
PhAsO (PAO) + Cl
+ 2H
2
PhAsO (PAO) + I
+ 2H
3
(Ph=phenyl)
®*
coated............................................................. 1 ....................... each.....................50085-00
®
Ampules ........................................ varies................ 16/pkg.................14268-10
Call
toll-free
800-227-4224
Contact the HACH office or distributor serving you.
www.hach.com
; E-mail –
O → PhAsO(OH)
+ 2Cl
+ 2H
2
2
O → 3I
+ PhAsO(OH)
+ 2H
2
2
Qty. required per test
techhelp@hach.com
+
+
Unit
HACH COMPANY
WORLD HEADQUARTERS
Telephone: (970) 669-3050
FAX: (970) 669-2932
50081_Chlorine_Total_Back.fm
Cat. No.
3/07 3ed

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents