Setting Titrant Concentration - Hach AutoCAT 9000 Instruction Manual

Chlorine amperometric titrator
Table of Contents

Advertisement

Software Setup

4.5 Setting Titrant Concentration

1.
From Titrants menu,
press
to select
1
Calibrate/Enter
Titrant Conc.
5.
Highlight User and
press the
CHECK MARK
to input or change the
user ID.
Software Setup
Page 40
14. Press
to confirm.
1
15. Select Rinse Before Replacing and follow the embedded method procedure.
2.
Highlight PAO 5.64mN
(or another method) and
press the
to
CHECK MARK
view the list of available
titrants for the chosen
procedure.
Note: 5.64 mN is equivalent to
0.00564 N.
6.
Use the alphanumeric
keypad to input the user ID.
Press the
to
CHECK MARK
validate.
Note: The user ID may be any
combination of alphanumeric
characters up to 16 characters
in length.
3.
Highlight the titrant
identifier (PAO 5.64mN in
this example) and press the
to validate.
CHECK MARK
Note: The AutoCAT 9000
automatically stores installed
titrants for later concentration
updates or calibrations.
7.
Verify the user ID and
press
to confirm.
1
4.
Press
to select Enter
1
Titrant Concentration.
8.
Highlight Titrant Conc.
and press the
CHECK MARK
to input or change the
concentration.